MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA AMISTEM H HA COATED LAT STEM SIZE 5; HIP CEMENTLESS STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA AMISTEM H HA COATED LAT STEM SIZE 5; HIP CEMENTLESS STEM Back to Search Results
Model Number 01.18.145
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 26-sep-2022: lot 133084: (b)(4) items manufactured and released on 15-oct-2013.Expiration date: 2018-08-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Clinical evaluation performed by medacta medical affairs manager: revision 8 years and 5 months of the femoral stem after primary tha due to stem loosening.No information concerning patient general health status and the presence of comorbidities is available.In the radiographic image provided, radiolucent lines and signs of stress shielding are visible.Aseptic loosening is a possible literature described adverse event after primary cementless hip arthroplasties and causes are often unknown.The reason of this failure cannot be determined.Picture analysis: looking at the image attached to the complaint it seems that almost the ha coating on the stem body has been correctly aboserbed by patient bone.It is not possible to determine the root cause of reported stem loosening.
 
Event or Problem Description
At about 8 years and 5 months after the primary surgery, the patient came in reporting pain due to a loose stem and the cause of the loose stem is unknown.The surgeon revised successfully the medacta stem and head with competitor components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMISTEM H HA COATED LAT STEM SIZE 5
Common Device Name
HIP CEMENTLESS STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15604268
Report Number3005180920-2022-00776
Device Sequence Number2315138
Product Code LZO
UDI-Device Identifier07630030804212
UDI-Public07630030804212
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K093944
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model Number01.18.145
Device Catalogue Number01.18.145
Device Lot Number133084
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/15/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-