| Model Number |
956-613 |
| Medical Device Problem Codes |
Corroded (1131); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the complaint the customer discovered a rusty needle on a safepico sampler.The customer returned the sampler box (100 pieces) back to radiometer on (b)(6) 2022.It was confirmed that the rusty needle was discarded and not used on a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Radiometer medical aps has concluded their investigation.Root cause of the needle corrosion has been identified as exposure of syringes to chlorine.By testing of reference samples, it has been concluded that exposure to chloride occurs after samplers have been shipped to china.Investigation of each step of the supply chain has been performed.Use of chloride containing gases has not been identified anywhere in the supply chain and would not be present during normal handling/storage/shipping.Hence, samplers from the same lot outside of china will not be affected.
|
| |
|