Model Number 68LP-275
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
Device analysis lab received a device with lot number 3272342 and catalog number 68lp-275.Healthcare professional reported via rga left side "any creases" and "implant totally deflated with some folding on itself".Device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: crease fold, wear abrasion, brown biological tissue inner the device and opening on posterior.Leak test and microscopic analysis was performed which identified: crease sharp on posterior and observed void 1 mm in non-textured, valve-to shell-bond area.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is:an opening crease sharp on posterior assessed as fold flaw opening.A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15605184
Report Number9617229-2021-59902
Device Sequence Number2215728
Product Code FWM
UDI-Device Identifier10888628000575
UDI-Public10888628000575
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/01/2022
Device Model Number68LP-275
Device Catalogue Number68LP-275
Device Lot Number3272342
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/14/2021
Initial Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-