Model Number 13827
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2022
Event Type  malfunction  
Event Description
Physician contacted technical solutions to report error codes 100 and 123 showing up on a patient's inquiry report.The physician reported that the patient noticed an error code 11 on their patient therapy controller.When the patient came in to have their pump inquired the next day, the inquiry report showed error codes 100 and 123.Technical solutions instructed the physician to perform an emergency pump stop, followed by a demand bolus of 0mg over one minute to reset the pump.When the pump was reinquired after these steps, no error codes were listed.The physician stated that the patient had not experienced any recent falls, accidents, or mris.The patient was not experiencing any symptoms.
 
Manufacturer Narrative
A review of the device history record, which includes verification of all steps in the manufacturing of the pump, verification of all final testing performed by/on the pump, verification of sterilization, and packaging for subject pump was performed.The review did not identify any non-conformances, issues or capas associated with pump function.Device remains implanted and was not returned for additional evaluation and investigation.Several attempts were made to follow-up with the physician's office to get additional information about this issue, but these attempts were unsuccessful.As physical investigation was not performed on the device, a definitive root cause could not be determined for the alleged issue.If additional information becomes available, a supplemental mdr will be sent.Internal complaint number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMETRA II PROGRAMMABLE PUMP
Type of Device
IMPLANTABLE INFUSION PUMP
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer (Section G)
FLOWONIX MEDICAL, INC.
120 forbes blvd
suite 170
mansfield MA 02048
Manufacturer Contact
james bennett
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key15605901
MDR Text Key306556934
Report Number3010079947-2022-00177
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00810335020228
UDI-Public(01)00810335020228(17)190804(10)25071
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/04/2019
Device Model Number13827
Device Catalogue Number13827
Device Lot Number25071
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/16/2022
Initial Date FDA Received10/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-