• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LDR MÉDICAL MOBI-C P&F IMPLANT HOLDER BODY; MOBI-C CERVICAL DISC PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MÉDICAL MOBI-C P&F IMPLANT HOLDER BODY; MOBI-C CERVICAL DISC PROSTHESIS Back to Search Results
Catalog Number MB9001R-2
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event or Problem Description
It was reported that the knob broke off a mobi-c inserter intra-operatively.There were no patient impacts.
 
Additional Manufacturer Narrative
Inspection: visual inspection revealed that the knob was fractured off of the end of the implant holder.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis forces applied during use.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Event or Problem Description
It was reported that the knob broke off a mobi-c inserter intra-operatively.There were no patient impacts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBI-C P&F IMPLANT HOLDER BODY
Common Device Name
MOBI-C CERVICAL DISC PROSTHESIS
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15606008
Report Number3004788213-2022-00081
Device Sequence Number16710121
Product Code MJO
UDI-Device Identifier03662663000741
UDI-Public(01)03662663000741(10)310232510/21
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P110009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMB9001R-2
Device Lot Number310232510/21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date03/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-