• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CHEM8+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CHEM8+ CARTRIDGE Back to Search Results
Catalog Number 09P31-26
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Hematoma (1884); Intracranial Hemorrhage (1891)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
On 27-sep-2022, abbott point of care was contacted by a customer regarding i-stat chem8+ cartridges that yielded suspected discrepant hematocrit result of 37% on a 17 month old female patient with an intracranial hemorrhage.There was no additional patient information available at the time of this report.(method; collect/tested; time; hgb(mmol/l); hct(% pcv); sample) i-stat 1, (b)(6) 2022, 02:04, 9.2, 37, a.Beckman, (b)(6) 2022, ni, 6.4, 18.8, a.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident #: (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Additional Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 10-nov-2022.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1 - product complaint level 2 and level 3 investigation procedure.Returned cartridge testing met the acceptance criterion for suppressed results but the sample size was too small (<17) to assess points outside total allowable error.No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CHEM8+ CARTRIDGE
Common Device Name
CHEM8+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key15606013
Report Number2245578-2022-00146
Device Sequence Number2263367
Product Code JGS
UDI-Device Identifier10054749005519
UDI-Public10054749005519
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K183688
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/26/2022
Device Catalogue Number09P31-26
Device Lot NumberH22150
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/27/2022
Supplement Date Received by Manufacturer11/10/2022
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age17 MO
Patient SexFemale
-
-