• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CNWTT3
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified the manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during the cataract surgery with intraocular lens (iol) implant procedure, found to be a scratch on optic of the iol.The physician removed the iol from patient eye.Additional information has been requested.
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.Iol received in unlabelled iol case.Iol in pieces in lens case base.A significant amount of solution is dried on the iol.Both haptics are bent.The optic is torn/split-cut dividing the iol in two and scratched/marked-rejectable with a piece missing.We are unable to determine the root cause for the reported complaint " a scratch was found on the optic ".The returned iol shows evidence of handling by the customer due to the presence of significant amount of solution is dried on the iol, the reported damage was highlighted post implantation.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol contributed to the event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15606065
Report Number9612169-2022-00531
Device Sequence Number16710771
Product Code MJP
UDI-Device Identifier00380652468251
UDI-Public00380652468251
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/11/2026
Device Model NumberCNWTT3
Device Catalogue NumberCNWTT3.060
Device Lot Number25366017
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/13/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-