| Model Number |
NSLX145C |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pc-(b)(4).Date sent:(b)(6)2022.Batch # unk.No lot or batch number was provided therefore a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a gyn procedure the jaws of the device locked on tissue and a hook had to be used to release.There were no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 11/3/2022 d4 batch #: v96909 investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All tones were heard during functional testing.No issues were noted with the opening and closing of the jaws.There were no anomalies noted with the functionality of the device.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device lot/batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|