Qn#(b)(4).The customer returned one unit 5-12615 et tube, cuffed, spiral-flex 7.5 for investigation.The returned et tube was visually examined with and without magnification.Visual examination of the returned et tube revealed that the sample appeared typical.No defects or anomalies were observed.A functional inspection was performed to attempt to inflate and deflate the balloon cuff on the returned et tube.A lab inventory syringe filled with air was connected to the valve of the pilot balloon.Upon injection of air, the pilot balloon was able to properly inflate.However, the balloon cuff did not inflate at all.The inflation line appeared to be blocked.After deflating the pilot balloon, water was injected into the inflation line to identify where the blockage in the inflation line was.Upon injection of the water, no water got past the distal end of the pilot balloon.Therefore, there appears to be an obstruction in the inflation line at the distal end of the pilot balloon.The obstruction was caused by adhesive material getting into the distal end of the pilot balloon during assembly and blocking the inflation line.A device history record review was performed and no relevant findings were identified.The ifu for this product states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is detected in any part of the inflation system, the tube should not be used." the reported complaint was confirmed based upon the sample received.Upon functional inspection, the inflation line was found to be obstructed at the distal end of the pilot balloon which prevented the balloon cuff from inflating properly.The obstruction was caused by adhesive material getting into the distal end of the pilot balloon during assembly and blocking the inflation line.A non-conformance was previously opened to further investigate this complaint issue by the manufacturing site.Corrective actions were implemented in aug 2022 to help prevent this issue from recurring.The returned sample was manufactured prior to these corrective actions.
|