• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Model Number VERSION 1.0
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a recanalization procedure, the wire allegedly fractured.It was further reported that the multiple attempts of retrieving the fractured wire were attempted but with no success.Reportedly, patient was sent to a hospital where ir physician was able to successfully retrieve fractured wire and patient was discharged.Additionally, snare and endovascular forceps were used at the center for the retrieval attempt.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: 07/2024).Device pending return.
 
Additional Manufacturer Narrative
The medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the catheter was not returned for evaluation.Only the guidewire was delivered and physically investigated.The guidewire was uncoiled at 1 cm from the tip of the guidewire.Therefore, the investigation is not confirmed for the reported break issue.However, the investigation is confirmed for material deformation issue.A clear root cause could not be identified but a broken guidewire represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiration date: 07/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a recanalization procedure, the wire allegedly fractured.It was further reported that the multiple attempts of retrieving the fractured wire were attempted but with no success.Reportedly, patient was sent to a hospital where ir physician was able to successfully retrieve fractured wire and patient was discharged.Additionally, snare and endovascular forceps were used at the center for the retrieval attempt.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Common Device Name
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15606883
Report Number3008439199-2022-00150
Device Sequence Number2263389
Product Code MCW
UDI-Device Identifier07640142810599
UDI-Public(01)07640142810599
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K211738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVERSION 1.0
Device Catalogue Number80238
Device Lot Number220512
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient RaceBlack Or African American
-
-