| Catalog Number |
07.02053.001 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a surgeon feels the vitality driver was not calibrated correctly, and did not output enough torque on the closure tops during surgery.There have been no reported patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that a surgeon feels the vitality driver was not calibrated correctly, and did not output enough torque on the closure tops during surgery.There have been no reported patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in h3.Additional information in h6: component, investigation type, findings, and conclusions.Inspection: the device was returned to the supplier for evaluation.The supplier tested the device and found that it was within calibration specifications.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined since the returned device was found to function as expected.Device usage this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|