• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II EXTENSION CUFF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II EXTENSION CUFF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number ETCF3232C49E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
An endurant iis bifurcate stent graft was implanted for the endovascular treatment of a 55mm abdominal aortic aneurysm.  it was reported that a ct carried out 6 months later identified a proximal type 1a endoleak without aneurysm growth.A 32 medtronic cuff and endoanchors were implanted on the same date, but a small leak remained.Per the physician the cause of the endoleak is  anatomy due to a conical neck.  no additional sequelae was reported and the patient will be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II EXTENSION CUFF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15607107
Report Number9612164-2022-03801
Device Sequence Number19411906
Product Code MIH
UDI-Device Identifier00613994991454
UDI-Public00613994991454
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/23/2023
Device Model NumberETCF3232C49E
Device Catalogue NumberETCF3232C49E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Initial Report FDA Received Date10/14/2022
Date Device Manufactured07/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexFemale
-
-