| Model Number |
15025-075-050 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The implant did not return for evaluation.A photograph was provided which confirmed the event.The identifying lot number were provided; therefore, a review of the device history record could not be conducted.The original surgery date is unknown.Based on the limited information provided, a root cause cannot be determined at this time.If additional information is provided, supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
A revision surgery was performed to extend a previously implanted construct to address degenerative disc disease.A screw adjuster was used to remove a screw.After removal, it was noted the load ring on the screw was deformed and dislocated.
|
| |
|