| Model Number |
FT18KN35NSC020N |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol b10009704, as a result of warning letter cms# 617147.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.The returned sample was reviewed.It was determined to be a device with a 50 mm shaft length.The part number identified on the labels was for a device with a 55 mm shaft length.The difference between was the curve of the shaft.The one does not have a curve; it was straight.The root cause of the reported issue was found to be a mix up occurred during labeling of the tray after the device was sealed in the package.
|
| |
|
Event or Problem Description
|
|
It was reported that two trachs were not the correct size.No patient injury was reported.
|
| |
|
Search Alerts/Recalls
|