DEPUY SYNTHES PRODUCTS LLC 10.5 CM QD ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC Back to Search Results
Model Number QD11-S-G1
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device is being further investigated.Once the investigation has been completed, a supplemental medwatch will be submitted.Udi: (b)(4).
 
Event or Problem Description
It was reported that during service and repair/pre-testing, it was observed that the motor device heated up at the elbow.The device also failed pretests for temperature assessment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2022.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device was heating up.The device also failed pretest for temperature assessment.Therefore, the reported condition was confirmed.The assignable root cause was traced to improper maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10.5 CM QD ANGLE ATTACHMENT
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key15607750
Report Number1045834-2022-01374
Device Sequence Number14003395
Product Code HBC
UDI-Device Identifier00845384016472
UDI-Public00845384016472
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQD11-S-G1
Device Catalogue NumberQD11-S-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/31/2015
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1