Model Number 3500-0600-82
Medical Device Problem Codes Defective Alarm (1014); Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr was generated under protocol (b)(4) as a result of warning letter cms#: 617147.Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.Visual inspection observed that the device was in good condition with no appearance of any physical damage.During the power up process, check and test force sensor testing, the reported problem was duplicated.The root cause of the reported issue was found to be the force sensor cable was damaged and pinched by right plunger case caused by the incorrect assembly by the customer.Replaced force sensor.Performed preventative maintenance and all functional testing which passed.
 
Event or Problem Description
It was reported that there was a system failure.Force password test, biomed passcode did not work.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDFUSION 3500 SYRINGE INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15608071
Report Number3012307300-2022-24135
Device Sequence Number2328263
Product Code FRN
UDI-Device Identifier10610586043017
UDI-Public10610586043017
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K890120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3500-0600-82
Device Catalogue Number3500-0600-82
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2022
Initial Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-