• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I 3I T3¿ SHORT EXTERNAL HEX PARALLEL WALLED IMPLANT, 5MM(D) X 5MM(L); DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3¿ SHORT EXTERNAL HEX PARALLEL WALLED IMPLANT, 5MM(D) X 5MM(L); DENTAL IMPLANT Back to Search Results
Model Number BOES505
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
Date of Event 08/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant was removed due to peri-implantitis with inflammation at the site.The doctor performed curettage.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).Patient id not provided.Patient age not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Dhr review was completed for the subject lot number 2015051511.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record op# 210 was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 2015051511 for similar events and no other complaint was identified.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3¿ SHORT EXTERNAL HEX PARALLEL WALLED IMPLANT, 5MM(D) X 5MM(L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15608136
Report Number0001038806-2022-01561
Device Sequence Number2136577
Product Code DZE
UDI-Device Identifier00844868033639
UDI-Public(01)00844868033639(17)200618(10)2015051511
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K150571
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/18/2020
Device Model NumberBOES505
Device Catalogue NumberBOES505
Device Lot Number2015051511
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer02/19/2023
03/07/2023
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date02/24/2023
03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/18/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight50 KG
-
-