| Model Number |
FT16AN40NSE073N |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tube, only used once and inspected prior to second insertion, was found to have a split flange hole during a pre-test.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.During the visual inspection of the tube, a tear through the entire eyelet was observed.The root cause of the reported issue could not determine root cause for the tear.A corrective and preventative action has been opened to address the reported issue.
|
| |
|