| Model Number |
518-036 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient''s date of birth, age unk.Patient''s gender unk.Patient''s weight unk.Patient''s ethnicity/race unk.Relevant tests/laboratory data unk.No concomitant devices reported.The device was not returned, thus no investigation could be completed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
A peripheral atherectomy procedure commenced to treat a lesion in the patient''s left anterior tibial (at) artery.The physician chose a spectranetics quick-cross support catheter to treat the patient.During use, the quick-cross device separated in the area of the lesion.The physician used a snare to successfully retrieve the separated portion from the at region.The procedure was completed with a new quick-cross catheter with no reported patient harm.This report captures the quick-cross for shaft separation, requiring intervention for removal.
|
| |
|
Event or Problem Description
|
|
This event is no longer reportable for shaft separation that required intervention to retrieve.Per the returned device analysis, the quick-cross catheter was returned intact with no separation observed.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: based on the device evaluation and investigation, this event is no longer reportable.D9: the device was returned to the manufacturer on 24oct2022.G3: the device evaluation and investigation were completed on 27oct2022.H3: the quick-cross device was returned intact with no separation observed, and showed normal signs of use (slightly crushed at the distal tip).However, a portion of a non-philips device was received along with the quick-cross.H6: based on the device evaluation and investigation, the event has been determined to have been caused by non-device related factors.Therefore, the heic code was corrected to 2199 (from 4641 on the initial mdr).Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|