Model Number 518-036
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient''s date of birth, age unk.Patient''s gender unk.Patient''s weight unk.Patient''s ethnicity/race unk.Relevant tests/laboratory data unk.No concomitant devices reported.The device was not returned, thus no investigation could be completed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
A peripheral atherectomy procedure commenced to treat a lesion in the patient''s left anterior tibial (at) artery.The physician chose a spectranetics quick-cross support catheter to treat the patient.During use, the quick-cross device separated in the area of the lesion.The physician used a snare to successfully retrieve the separated portion from the at region.The procedure was completed with a new quick-cross catheter with no reported patient harm.This report captures the quick-cross for shaft separation, requiring intervention for removal.
 
Event or Problem Description
This event is no longer reportable for shaft separation that required intervention to retrieve.Per the returned device analysis, the quick-cross catheter was returned intact with no separation observed.
 
Additional Manufacturer Narrative
B5: based on the device evaluation and investigation, this event is no longer reportable.D9: the device was returned to the manufacturer on 24oct2022.G3: the device evaluation and investigation were completed on 27oct2022.H3: the quick-cross device was returned intact with no separation observed, and showed normal signs of use (slightly crushed at the distal tip).However, a portion of a non-philips device was received along with the quick-cross.H6: based on the device evaluation and investigation, the event has been determined to have been caused by non-device related factors.Therefore, the heic code was corrected to 2199 (from 4641 on the initial mdr).Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS QUICK-CROSS SUPPORT CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key15609424
Report Number1721279-2022-00176
Device Sequence Number17934413
Product Code DQY
UDI-Device Identifier00813132020675
UDI-Public(01)00813132020675(17)221016(10)FQP20K15A
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K033678
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/16/2022
Device Model Number518-036
Device Catalogue Number518-036
Device Lot NumberFQP20K15A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer10/27/2022
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER MANUFACTURER/SIZE UNK; GUIDE WIRE, MANUFACTURER/SIZE UNK; INTRODUCER SHEATH MANUFACTURER/SIZE UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-