It was reported that after 2 weeks of a correction of left patellar dislocation under general anesthesia performed on (b)(6) 2022, using biosure, there was fluid coming out from the wound.The wound healed after debridement performed on (b)(6) 2022.No further complications were reported.
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no relationship found between the device and the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.It was communicated that the wound healed after the debridement was performed.Should any additional clinical information be provided, this complaint will be re-evaluated.The complaint was not confirmed.Factors that could have contributed to the reported event include insufficient or improper cleaning of the surgical site, or a biological response unrelated to the device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|