Model Number 74123148
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a thr surgery was performed on (b)(6) 2016, the patient experienced migration and collapse of a bhr resurfacing femoral head 48mm due to bad bone quality.This adverse event was treated by a revision surgery on (b)(6) 2022, in which a the femoral head was replaced with a femoral stem and a dual mobility insert.The patient's current health status is unknown.
 
Additional Manufacturer Narrative
It was reported that a revision surgery was performed due to migration and collapse of a bhr femoral head related to unhealthy bone tissue.The patient's current health status is unknown.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the device was performed using the batch number, part number and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for he femoral head, this will continue to be monitored.In the absence of the actual device, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the product and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.The available medical documents were reviewed.The reported ¿bad bone quality¿ cannot be ruled out as a contributing factor to the reported migration and collapse of the femoral head and subsequent revision 6 years later.The patient impact beyond the revision and expected convalescence period cannot be determined with the limited information provided.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Besides the unhealthy bone tissue, other specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement.Should the device or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Event or Problem Description
It was reported that, after a thr surgery was performed on (b)(6) 2016, the patient experienced migration and collapse of a bhr resurfacing femoral head 48mm due to unhealthy bone tissue.This adverse event was treated by a revision surgery on (b)(6) 2022, in which a the femoral head was replaced with a femoral stem and a dual mobility insert.The patient's current health status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BHR RESURFACING FEMORAL HEAD 48MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15610824
Report Number3005975929-2022-00476
Device Sequence Number16825553
Product Code NXT
UDI-Device Identifier03596010552426
UDI-Public03596010552426
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/23/2021
Device Model Number74123148
Device Catalogue Number74123148
Device Lot Number16AW03994
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer09/20/2022
11/03/2022
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date10/24/2022
11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/25/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
Patient Weight64 KG
-
-