• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-55
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 10/07/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that , the cardiosave intra-aortic balloon pump (iabp) unit had many failures; faulty x3 and x2 sensor; leak drive; and pim issues.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and was able to reproduce the reported issue.The fse found error codes 58 pneumcntr system failure -shutdown, 124 pneumcntr prolonged shuttle prs system failure and 37 fill fail-shuttle purge timeout in the log.Following a problem with blood flow, the fse noticed that there was blood in the circuit so the fse changed the helium reservoir assy, drive manifold assembly, pneumatic module assy and mufflers that were impacted by the blood to resolve the issue.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15610899
MDR Text Key306076457
Report Number2249723-2022-02669
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/07/2022
Initial Date FDA Received10/14/2022
Supplement Dates Manufacturer Received11/16/2023
Supplement Dates FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-