SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SETS - FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7386-24
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the line was leaking where the tubing inserts into the luer end.Second unit reported similar problems with the same lot number.Replaced with new lot and no reported issues.There was no patient injury reported.
 
Additional Manufacturer Narrative
Operator of device is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: updated.Device evaluation: the samples consist of nine units; the returned samples were received in unused conditions within original package.The sample was visually inspected under normal conditions of illumination to detect any condition that could cause the failure mode reported.The samples don?t present any damage that could cause the failure mode reported.The sample received was set for leak test using a hydrostat vessel.No leaks were detected on nine units; complaint was not confirmed.No other analysis was performed.Root cause cannot be addressed since no defects were found in samples received for evaluation.No action required since the complaint was not confirmed.No fault was found.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SETS - FLOW STOP
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
unit #8
minneapolis, MN 55442
MDR Report Key15610983
Report Number3012307300-2022-24217
Device Sequence Number7169671
Product Code FPA
UDI-Device Identifier15019517161977
UDI-Public15019517161977
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K031361
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number21-7386-24
Device Catalogue Number21-7386-24
Device Lot Number4254183
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date06/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-