• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RENEGADE; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RENEGADE; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 29417
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the catheter was broken.A 105/3.0/2.5/.021/20/straight renegade microcatheter was selected for use in the liver.During the procedure, it was noted that the catheter was broken.The procedure was completed with another of the same device.No complications were reported and patient is good post procedure.
 
Event or Problem Description
It was reported that the catheter was broken.A 105/3.0/2.5/.021/20/straight renegade microcatheter was selected for use in the liver.During the procedure, it was noted that the catheter was broken.The procedure was completed with another of the same device.No complications were reported and patient is good post procedure.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.The hub and shaft were visually inspected when returned.The renegade device showed multiple small bends and kinks.The shaft showed stretching and a fracture located 33cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15611035
Report Number2124215-2022-39631
Device Sequence Number15248738
Product Code DQO
UDI-Device Identifier08714729375395
UDI-Public08714729375395
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K020012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2025
Device Model Number29417
Device Catalogue Number29417
Device Lot Number0028983638
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-