| Model Number |
29417 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the catheter was broken.A 105/3.0/2.5/.021/20/straight renegade microcatheter was selected for use in the liver.During the procedure, it was noted that the catheter was broken.The procedure was completed with another of the same device.No complications were reported and patient is good post procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter was broken.A 105/3.0/2.5/.021/20/straight renegade microcatheter was selected for use in the liver.During the procedure, it was noted that the catheter was broken.The procedure was completed with another of the same device.No complications were reported and patient is good post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the complaint device was received for product analysis.The hub and shaft were visually inspected when returned.The renegade device showed multiple small bends and kinks.The shaft showed stretching and a fracture located 33cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
| |
|