• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE VERSATIE SYSTEM; BONE FIXATION CERCLAGE, SUBLAMINAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE VERSATIE SYSTEM; BONE FIXATION CERCLAGE, SUBLAMINAR Back to Search Results
Model Number 86101245
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 09/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product was not returned and no images were provided to confirm the complaint.The root cause could not be determined though review of the reported event suggests a possible allergic reaction or possible band movement/ tissue irritation.Should more information become available a follow-up report will be completed.Labeling review: "potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include: early or late infection.Potential risks identified with the use of this system, which may require additional surgery, include: infection, foreign body reaction to the implants." "warnings, cautions and precautions: implants can loosen, fracture, corrode, migrate, possibly increase the risk of infection, cause pain, or stress shield bone even after fracture has healed.If a device remains implanted after complete healing, it can actually increase the risk of refracture in an active individual.Therefore, these devices are temporary and should be removed after completing their intended function i.E., aiding in the bone healing process.The surgeon should weigh the risk versus benefit when deciding whether to remove an implant.Implant removal should be followed by adequate postoperative management to avoid refracture.If the patient is older, and has a low activity level, the surgeon may choose not to remove the implant, thus eliminating the risks associated with a second surgery.Care should be taken to insure that all components are ideally fixated prior to closure."pre-operative warnings: care should be used during surgical procedures to prevent damage to the devices and injury to the patient.".
 
Event or Problem Description
It was reported that approximately 6-8 months after implantation of versatie product, the patient developed an inflammatory reaction and possible unknown deficit.A removal procedure was performed where the versatie was removed and the patient continued with fusion with possible extension of construct.During removal, it was noted that the inflamed tissue was adjacent to the versatie tether.Per the reporter, there were no issues during initial implantation and surgical technique was followed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE VERSATIE SYSTEM
Common Device Name
BONE FIXATION CERCLAGE, SUBLAMINAR
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key15611063
Report Number2031966-2022-00204
Device Sequence Number2749581
Product Code OWI
UDI-Device Identifier00887517984296
UDI-Public887517984296
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K161265
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number86101245
Device Lot Number220265
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Initial Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexPrefer Not To Disclose
-
-