• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS Back to Search Results
Model Number 7312
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code Syncope/Fainting (4411)
Date of Event 09/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported that he performed blood glucose testing with the contour next link 2.4 meter and the reading obtained on the meter was higher than that on the insulin pump.No specific readings were provided.The insulin pump delivered insulin based on the incorrect reading and the customer fainted.The customer stated that the event occurred three times and every time he had to call an emergency service.The customer stated that the physician adjusted the insulin pump to deliver right amount of insulin.No further event details were provided.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
As per contour next link 2.4 blood glucose monitoring system user guide, "wash hands and dry well before handling to keep the meter and test strips free of water, oils and other contaminants.Please follow all instructions for use and care exactly as described to help avoid inaccurate results." the customer did not wash his hands prior to testing.The patient/family was the initial reporter, so personal information was not entered.No information was captured in section a4 as the customer's weight was not provided.
 
Additional Manufacturer Narrative
The customer did not return the device for evaluation.Therefore, an in-house testing was performed using blood sample with the in-house contour next link 2.4 meter and in-house contour next test strips from lot#: 1epeg11a, which gave a satisfactory performance.Additionally, a device history record was reviewed for the suspected contour next link 2.4 meter and no manufacturing anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE TEST STRIPS
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION NUMBER: 1000134386)
2131-1 minamigata
toon ehime, 791-0 395
JA   791-0395
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key15611141
Report Number1810909-2022-00160
Device Sequence Number2761911
Product Code NBW
UDI-Device Identifier10301937312219
UDI-Public10301937312219
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P150001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/31/2023
Device Model Number7312
Device Catalogue Number7312
Device Lot Number1EPEG11A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date01/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-