| Model Number |
1012453-12 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly calcified, moderately tortuous, 90% stenosed lesion in the left anterior descending (lad) artery.A 4x12mm nc trek balloon dilatation catheter (bdc) was advanced to the target lesion and upon the initial inflation, a small leak in the balloon was observed.The bdc was immediately removed from the anatomy and another device was used to successfully post-dilatate the lesion.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for analysis.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual and functional inspections were performed on the returned device.The reported leak was confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported leak or noted balloon rupture on the returned device.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|