• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD VENFLON¿ IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD VENFLON¿ IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 391592
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  malfunction  
Manufacturer Narrative
The manufacturing location for this product is (b)(4), india.This site is not registered with the fda.Therefore, bd corporate headquarters in (b)(4) has been listed in manufacturer name, city and state and mfr site and the (b)(4) fda registration number has been used for the manufacture report number.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that 2 bd venflon¿ iv catheter tips were defective.The following information was provided by the initial reporter: "complain on venflon i catheter tip hardness".
 
Manufacturer Narrative
Follow up mdr for device evaluation: there are no samples and no photograph along with the reported complaint of catheter damage/defective.The investigating team has used the retention samples of material number 391592 and lot number 2160678 for investigating the reported defect.The investigation and simulation were carried out on two retention samples.The investigating team has visually tested the retention samples for catheter damage/defective and no catheter damage/defective was found in the two retention samples.Since there is no samples and no photograph, the defect could not be confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VENFLON¿ IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15611393
MDR Text Key307109057
Report Number2243072-2022-01766
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903915927
UDI-Public(01)00382903915927
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number391592
Device Lot Number2160678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/20/2022
Initial Date FDA Received10/14/2022
Supplement Dates Manufacturer Received10/26/2022
Supplement Dates FDA Received07/01/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-