| Model Number |
3500-0600-82 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A manufacturing device history record (dhr) review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Both tamper seals were intact.Right case was found cracked.Auxiliary control unit watchdog failure and primary audible alarm post.Was found in pump event history log.Reported problem was unable to be duplicated during power-up test.The root cause of the reported issue was unable to be determined.Actions were taken to mitigate the reported issue: replaced interconnect board and primary speaker as a preventive measure.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was presenting with auxiliary control unit watchdog failure and primary audible alarm post.No patient injury was reported.
|
| |
|
Search Alerts/Recalls
|