• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941820350
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the shaft break occurred.The target lesion was severely tortuous and moderately to severely calcified mid to distal right coronary artery.Pre-dilatation was performed and a 3.50 x 20mm synergy xd drug-eluting stent was advanced for treatment however, during delivery of the device to the lesion, the physician was pushing with a lot of force and the shaft of the stent broke near the hub.The physician pulled the stent shaft out and the procedure was completed with a new synergy stent.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: synergy xd 3.50 x 20mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The stent outer diameter was measured and the result was within max crimped stent profile measurement the balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along the length of the hypotube shaft and a break 22 cm distal to distal end of strain relief.A visual and tactile examination of the outer and mid-shaft section found no issues.
 
Event or Problem Description
It was reported that the shaft break occurred.The target lesion was severely tortuous and moderately to severely calcified mid to distal right coronary artery.Pre-dilatation was performed and a 3.50 x 20mm synergy xd drug-eluting stent was advanced for treatment however, during delivery of the device to the lesion, the physician was pushing with a lot of force and the shaft of the stent broke near the hub.The physician pulled the stent shaft out and the procedure was completed with a new synergy stent.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15612391
Report Number2124215-2022-41405
Device Sequence Number20627643
Product Code NIQ
UDI-Device Identifier08714729981053
UDI-Public08714729981053
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/17/2023
Device Model NumberH7493941820350
Device Lot Number0028651014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date10/15/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-