| Model Number |
0500318N |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A user facility registered nurse (rn) reported that an internal dialyzer blood leak occurred five minutes after the start of a patient¿s hemodialysis (hd) treatment.The rn confirmed no external blood leak was observed.The machine, a fresenius 2008t, alarmed appropriately with a minor blood leak alert.Fresenius combi set bloodlines were being utilized for the treatment.Blood test strips were used and tested for the presence of blood.No damage was identified on the dialyzer prior to use.Per rn the patient¿s blood was not returned and stated the estimated blood loss (ebl) was 300 ml.Immediately following the event, the patient was re-setup with new supplies on a different machine where they were able to complete treatment.It was confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The dialyzer was discarded and was available to be returned for manufacturer evaluation.
|
| |
|
Event or Problem Description
|
|
A user facility registered nurse (rn) reported that an internal dialyzer blood leak occurred five minutes after the start of a patient¿s hemodialysis (hd) treatment.The rn confirmed no external blood leak was observed.The machine, a fresenius 2008t, alarmed appropriately with a minor blood leak alert.Fresenius combi set bloodlines were being utilized for the treatment.Blood test strips were used and tested for the presence of blood.No damage was identified on the dialyzer prior to use.Per rn the patient¿s blood was not returned and stated the estimated blood loss (ebl) was 300 ml.Immediately following the event, the patient was re-setup with new supplies on a different machine where they were able to complete treatment.It was confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The dialyzer was discarded and was available to be returned for manufacturer evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3 plant investigation: complaint device returned to the manufacturer for physical evaluation.Sample subjected to bubble point test and no leak detected.Possibly due to the coagulated blood in the headers.Dialyzer subjected to destructive disassembly for further visual examination.Upon extraction of the fiber bundle, a potted fiber fragment was identified at approximately 200°, with dialysate ports situated at 0°, on non-cavity id end.Fiber fragment was approximately 2.0 inches in length and was diagonally positioned along outer perimeter of the pu, away from the fiber bundle.An opposing end could not be isolated.Further examination of fiber bundle did not reveal any other damage or irregularities.No cracks noted on the dialyzer and only one fiber fragment was identified (nurse reported a crack and multiple fibers were moving around in the dialyzer).Lot history review performed.No other complaints reported against the lot.Production records review performed on reported lot.Investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.The investigation into the complaint was able to confirm the reported event.There is no recall associated with this complaint.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas: 2018-0171 (vision systems) and 2018-0161 (blood leak reduction) are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
| |
|