Model Number 690-00-28F
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
 
Event or Problem Description
It was reported that during a revision of the patient's right hip,, the constrained liner was impacted and reduced.After reduction, the inner bipolar portion of the liner dissociated from the outer portion.The liner was removed and another liner of a different lot was used.
 
Event or Problem Description
It was reported that during a revision of the patient's right hip,, the constrained liner was impacted and reduced.After reduction, the inner bipolar portion of the liner dissociated from the outer portion.The liner was removed and another liner of a different lot was used.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation involving a trident liner was reported.The event was confirmed.Method & results product evaluation and results: the constrained liner including the uhr bipolar head was returned for evaluation.The uhr bipolar head is disassociated from the poly liner.Visual inspection of the returned device indicated damage resulting from explantation.Clinician review: no medical records were received for review with a clinical consultant.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: it was reported that the patient was revised due to dislocation whereby during surgery the constrained liner failed.The constrained liner including the uhr bipolar head was returned for evaluation.The uhr bipolar head is disassociated from the poly liner.Visual inspection of the returned device indicated damage resulting from explantation.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG CONSTRAINED INSERT 28F
Common Device Name
PROSTHESIS, HIP, CONSTRAINED, CEMENTED OR UNCEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key15612885
Report Number0002249697-2022-01492
Device Sequence Number7148872
Product Code KWZ
UDI-Device Identifier07613327025552
UDI-Public07613327025552
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K061654
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model Number690-00-28F
Device Catalogue Number690-00-28F
Device Lot NumberVA7VK5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date10/16/2022
Supplement Report FDA Received Date01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
Patient SexFemale
-
-