• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2272
Device Problems Over-Sensing (1438); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
Related manufacture reference number: 2017865-2022-41860.It was reported that the patient presented prior to procedure.Upon interrogation, it was noted that the pacemaker and right ventricular lead exhibited oversensing and high capture threshold.The pacemaker was explanted and the lead capped on (b)(6) 2022.Both products were replaced during the same procedure.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of oversensing, and capturing issue were not confirmed.As received a device was returned above elective replacement level (eri) for analysis.Electrical and mechanical analysis, sensitivity tests, and visual inspection did not identify any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15613004
MDR Text Key301815819
Report Number2017865-2022-41859
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509589
UDI-Public05414734509589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model NumberPM2272
Device Catalogue NumberPM2272
Device Lot NumberA000059302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/27/2022
Initial Date FDA Received10/16/2022
Supplement Dates Manufacturer Received12/03/2022
Supplement Dates FDA Received12/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL LEAD
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexFemale
-
-