BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION Back to Search Results
Model Number 448100
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: na.(b)(6).Pma/510(k) #: k020321, k040099, k131331.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using the bd phoenix¿ automated microbiology system there was an error id result showing.No patient impact was reported.The following information was provided by the initial reporter: error id result showing.
 
Event or Problem Description
It was reported that while using the bd phoenix¿ automated microbiology system there was an error id result showing.No patient impact was reported.The following information was provided by the initial reporter: error id result showing.
 
Additional Manufacturer Narrative
H.6 investigation summary: a failure was reported on a phoenix 100 instrument (p/n 448100, s/n (b)(6)).A field service engineer (fse) was dispatched and proceeded to clean the instrument.Once the fse performed the cleaning the issue was resolved.The root cause is not able to be determined.This is an unconfirmed failure of a bd product.Samples were not received by quality for investigation and thus, returned material investigation could not occur.If samples are received at a later date, the complaint may be reopened.This complaint is not an early life failure (elf) of the device.Since the installation of this instrument, there have been service events, including pms (preventative maintenance) which have altered the instrument from its original state.Therefore, review of the device history record (dhr) from manufacturing is not applicable.Service history review for this instrument was completed and revealed no previous complaints related to results issues.Bd quality will continue to closely monitor for trends associated with this complaint.No new risks or hazards, or changes to existing risks/hazards, were identified as a result of this complaint.Per baltrmphxidastinstraph rev 6 the potential risk of the reported failure mode is user annoyance, which is an s1; reference row id 9.12.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM
Common Device Name
SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15613207
Report Number1119779-2022-01304
Device Sequence Number5810771
Product Code LON
UDI-Device Identifier00382904481001
UDI-Public00382904481001
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
SEE H.10
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number448100
Device Catalogue Number448100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date12/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/22/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-