|
A physician was attempting to use a visipro stent for treatment of a 30 mm calcified and fibrous lesion with 80% stenosis in the mid left iliac artery.The artery had a diameter of 7mm with mild tortuosity and moderate calcification.The device was prepped as per the ifu with no issues identified.It was reported the stent wouldn¿t track over the wire to the intended stent placement site, the physician removed the stent and balloon undeployed. the physician then pre dilated with a 6 mm balloon.When trying to re insert the stent, it looked kinked up on the side and requested a new stent that wasn¿t damaged. the second stent tracked and was placed without difficulty.No patient injury was reported.
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|