| Model Number |
AF5140MB |
| Medical Device Problem Codes |
Defective Component (2292); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The duckbilled valve wouldn't open during a code.
|
| |
|
Additional Manufacturer Narrative
|
|
A duckbill not opening during code would lower the volume delivered to the patient.Based on the reported information, the criteria for reporting an adverse event have been met.
|
| |
|
Event or Problem Description
|
|
The duckbilled valve wouldn't open during a code.
|
| |
|
Additional Manufacturer Narrative
|
|
A duckbill not opening during code would lower the volume delivered to the patient.Based on the reported information, the criteria for reporting an adverse event have been met summary: returned product confirmed the complaint.During investigation of the device without a manufacturing date/lot number, it was found that the device was made prior to certain engineering changes dating to 2015.Ra: rma-20020 contains appropriate risk assessment.Previous corrective actions of capa14-016 and mi-co19-0225 address the non-conformance identified by this complaint.
|
| |
|