MAUDE Adverse Event Report: CARDINAL HEALTH THORA SEAL 3; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH THORA SEAL 3; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number 8884713308
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Low Oxygen Saturation (2477)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the hospital used chest bottle with obvious creases in the drainage hose of the chest bottle, causes the drainage to be unsmooth, causing the patient¿s blood oxygen to drop rapidly and still under treatment at the hospital.Several attempts to gather information from the customer were made.To date, no response has been received.If additional pertinent information becomes available, the report will be updated.
 
Additional Manufacturer Narrative
The lot number was provided, and the device history record (dhr) was reviewed indicating that the product was released accomplishing all quality standards.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A physical sample was not received for the investigation.However, a picture was provided for the analysis.Upon reviewing the picture, kink was observed and has been covered by a white tape, this tape is not placed on the tube at the manufacturing site.From the review of the picture provided, the reported condition is confirmed.The probable root cause of the reported issue could be due to the tube was wrapped too tightly causing the kink.As per standard work instruction, coil the patient drainage tube in a counterclockwise circle, ensuring no kinks in the tube, standardize the method of coiling and securing the tube to prevent the tube from kinking as a kinked tube may prevent the unit from functioning.A quality alert was issued to the manufactured line to make them aware of the reported issue.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORA SEAL 3
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
sragh industrial estate, co, t
offaly
EI  
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15615102
Report Number9611018-2022-00565
Device Sequence Number17632633
Product Code KDQ
UDI-Device Identifier20884521101118
UDI-Public20884521101118
Combination Product (Y/N)N
Initial Reporter CountryTW
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884713308
Device Catalogue Number8884713308
Device Lot Number21K170FHX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer09/30/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-