• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. WISP MASK; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. WISP MASK; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 1094051
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Arrhythmia (1721); Dizziness (2194)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer received information that an end user alleged that she experienced lightheadedness, dizziness, an irregular heartbeat, and changes to heart rate at night after wearing the wisp mask with magnetic connections.The user had pacemaker placed in (b)(6) 2022 and has a monitor in her room that she uses while sleeping.The monitor has allegedly alerted her to dramatic changes in heart rate at night.There was no report of medical intervention.There was no report of serious patient harm or injury.The manufacturer gave information on the plastic clips for the wisp mask and instructed the user to stop using the mask until she contacts her cardiologist and dme for plastic clips.No product will be returned for investigation.The wisp instructions for use includes the following warnings: the headgear clips contain magnets.Contact your healthcare professional before you use this mask.Some medical devices may be affected by magnetic fields.The magnetic clips in this mask should be kept at least 2 in.(50 mm) away from any active medical device with special attention to implanted devices such as pacemakers, defibrillators and cochlear implants.Do not use in or near magnetic resonance imaging (mri) equipment.Keep unassembled magnetic headgear clips out of reach of children.In case of accidental swallowing, seek medical assistance immediately.The manufacturer will continue to monitor complaints for similar issues.The manufacturer concludes no further action is needed at this time.
 
Additional Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging visualization of particles related to a cpap device's sound abatement foam.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Section h6 updated in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WISP MASK
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key15615362
Report Number2518422-2022-86392
Device Sequence Number5983327
Product Code BZD
UDI-Device Identifier00606959006829
UDI-Public00606959006829
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K121631
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1094051
Device Catalogue Number1094051
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexFemale
-
-