Model Number 9.5 MM FLIP CUTTER II
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an anterior cruciate ligament all inside surgery a part of the device broke off without improper handling during tibial drilling in the patient's knee.All broken parts were retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual evaluation, it was noted the cutter tip is broken, and one of the distal tips of the shaft is bent.The fragment detailed in the event description was not returned for investigation.The observed condition is most likely the result of leveraging the device during use.The cause remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9.5 MM FLIP CUTTER II
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15615590
Report Number1220246-2022-05634
Device Sequence Number12230741
Product Code HWE
UDI-Device Identifier00888867004276
UDI-Public00888867004276
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number9.5 MM FLIP CUTTER II
Device Catalogue NumberAR-1204AF-95
Device Lot Number0024100684
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date02/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-