| Model Number |
392544 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of birth: only the patient's age was provided, therefore a default date of birth has been listed fda notified: the initial reporter notified the fda via uf/importer report#: (b)(4).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of bd insyte¿ autoguard¿ bcs experienced needles that would not retract.The following information was provided by the initial reporter: nurse was inserting 18 g bd iv catheter, when pressed button to retract the needle, needle wouldn't retract.Multiple instances that the needle would not retract as designed.Did you experience any adverse events as a result of this reported defect? no.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary since no samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.A device history record review showed no non-conformances associated with this issue during the production of this batch.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of bd insyte¿ autoguard¿ bcs experienced needles that would not retract.The following information was provided by the initial reporter: nurse was inserting 18 g bd iv catheter, when pressed button to retract the needle, needle wouldn't retract.Multiple instances that the needle would not retract as designed.Did you experience any adverse events as a result of this reported defect? no.
|
| |
|
Search Alerts/Recalls
|