• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW COVID-19 TEST KIT 24T (EUA); REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW COVID-19 TEST KIT 24T (EUA); REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID Back to Search Results
Catalog Number 190-000
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The remainder of the investigation remains in progress.A supplemental report will be provided after completion.Device discarded, single-use.
 
Event or Problem Description
The customer reported a false negative result with the id now covid-19 assay performed on (b)(6) 2022 using a nasal swab.Confirmation testing (platform: core lab) was performed on a nasopharyngeal swab in viral transport media and generated a negative result the same day.The customer stated that no further information is available.
 
Additional Manufacturer Narrative
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 1086021 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 190-000 / lot: 1086021, test base part number 190-430 / lot: 1086021.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 1086021 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough, inc.Was unable to determine the exact root cause of the reported issue.H3 other text : device discarded, single-use.
 
Event or Problem Description
The customer reported a false negative result with the id now covid-19 assay performed on (b)(6) 2022 using a nasal swab.Confirmation testing (platform: core lab) was performed on a nasopharyngeal swab in viral transport media and generated a negative result the same day.The customer stated that no further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ID NOW COVID-19 TEST KIT 24T (EUA)
Common Device Name
REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
6613888803
MDR Report Key15615671
Report Number1221359-2022-10009
Device Sequence Number20835038
Product Code QJR
UDI-Device Identifier10811877011269
UDI-Public10811877011269
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
EUA2000074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/09/2023
Device Catalogue Number190-000
Device Lot Number1086021
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer01/29/2023
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-