| Model Number |
622275 |
| Medical Device Problem Codes |
Degraded (1153); Suction Problem (2170)
|
| Health Effect - Clinical Code |
Pneumothorax (2012)
|
| Date of Event |
09/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: event description: provider was performing a centesis and when bottle had about 300ml left to fill a hissing sound was heard and air began to fill patient lung.The grey rubber top was cracked and broke thus breaking the vacumm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The complaint samples and two (2) pictures were submitted to the manufacturer for evaluation.Through visual inspection of the photographs and samples, it was observed the grey drain connector is cracked/ripped.A check of the production process shows that such a fault cannot occur in the manufacture of the product.The crack on the grey drain connector could be caused by the combined effect of gamma sterilization and uv radiation, which could have resulted from improper storage.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|