Model Number SN60AT
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A health care professional returned a product with a description of unspecified defect.Additional information has been received and stated that, the surgery was delayed for 10 minutes and completed by using another iol.There was no patient harm reported.
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.Iol (intraocular lens) returned positioned incorrectly in the iol case.A significant amount of solution is dried on the iol.One haptic is bent/deformed, the other is intact.The optic is cracked/fractured -rejectable.The complainant indicates the use of non company ovd (ophthalmic viscosurgical device) as a viscoelastic , which is not qualified to be used associated iol model.While we are unable to determine the origin of the reported complaint, our observations reasonably suggest that it is not manufacturing related.Based on the information provided by the customer, the most likely root cause is failure to follow ifu (instructions for use), as the surgeon states the use of non-qualified viscoelastic.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15615988
Report Number9612169-2022-00535
Device Sequence Number12784091
Product Code HQL
UDI-Device Identifier00380650130372
UDI-Public00380650130372
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN60AT
Device Catalogue NumberSN60AT.235
Device Lot Number25375768
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date12/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEALON.; MONARCH III IOL CARTRIDGE C.