• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH UNWTD KANG FEED TUBE 12FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH UNWTD KANG FEED TUBE 12FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884711501
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that a doctor inserted a new feeding tube into a patient but was unable to aspirate gastric content from the side port.The tube was then removed from the patient and the users tried to flush both ports and realize that the side port is occluded, neither air nor water could be flushed through the side port.A new feeding tube from the same lot was opened and inserted into same patient and both ports in the newly opened tube was working well.No patient injury.
 
Additional Manufacturer Narrative
A sample was not returned for evaluation.The device history record (dhr) file was reviewed, and no discrepancy was found according to the reported issue.The manufacturing process was reviewed.All processes and controls were found properly followed, including sub-assemblies, finished product assembly, packaging and inspections performed to the product.There were no abnormal conditions found that could trigger the reported condition.Furthermore since no samples or photos were returned, the reported issue could not be confirmed, and a possible root cause cannot be determined.If a sample is returned at a later date the investigation will be updated accordingly.Based on the available information, an action plan is not required at this time.The complaint will be used for tracking and trending purposes.
 
Additional Manufacturer Narrative
A sample was not returned for evaluation.The device history record (dhr) file was reviewed, and no discrepancy was found according to the reported issue.The manufacturing process was also reviewed.All processes and controls were found properly followed, including sub-assemblies, finished product assembly, packaging and inspections performed to the product.There were no abnormal conditions found that could trigger the reported condition.Based on the available information and since no samples or photos were returned, the reported issue could not be confirmed, and a possible root cause cannot be determined.If a sample is returned at a later date, the investigation will be updated accordingly.No action plan is required at this time.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNWTD KANG FEED TUBE 12FR
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15616016
Report Number9612030-2022-03423
Device Sequence Number20818663
Product Code KNT
UDI-Device Identifier10884521009684
UDI-Public10884521009684
Combination Product (Y/N)N
Initial Reporter CountrySN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8884711501
Device Catalogue Number8884711501
Device Lot Number1933040064
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer10/10/2022
10/10/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date11/04/2022
03/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-