| Model Number |
X SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device's screen turned completely white and was illegible.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.The customer's report was duplicated and attributed to the lcd display panel.The lcd panel was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|