CETERIX ORTHOPAEDICS, INC NOVOSTITCH PRO MENISCAL RPR SYS 0; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number CTX-A004
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during an arthroscopy, the upper jaw of the novostitch pro meniscal repair system broke, the broken piece was successfully removes by grasper.The procedure was completed using a smith and nephew back-up device.They wanted a result with two luggage tag sutures but needed to settle with one passing of each suture.Non-significant surgical delay or further complications were reported.
 
Event or Problem Description
It was reported that, during an arthroscopy, the upper jaw of the novostitch pro meniscal repair system broke, the broken piece was successfully removes by grasper.The procedure was completed using a smith and nephew back-up device.They wanted a result with two luggage tag sutures but needed to settle with one passing of each suture.Non-significant surgical delay or further complications were reported.The current health status of the patient is good.
 
Additional Manufacturer Narrative
B5 updated.
 
Additional Manufacturer Narrative
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device tip, attempted correction of a damaged device, or an inadvertent impact event.Per case details, the broken jaw was retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual inspection of the returned device found that it is not in its original packaging.The device was returned with the articulation bar for the top jaw fractured.There is biological debris on the device and the lower jaw is intact.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Per case details, the broken jaw was retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with component failure.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device tip, attempted correction of a damaged device, or an inadvertent impact event.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVOSTITCH PRO MENISCAL RPR SYS 0
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
CETERIX ORTHOPAEDICS, INC
6500 kaiser dr ste 120
fremont CA 94555
Manufacturer (Section G)
CETERIX ORTHOPAEDICS, INC
6500 kaiser dr ste 120
fremont CA 94555
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15616412
Report Number3009131204-2022-00022
Device Sequence Number5912009
Product Code GAT
UDI-Device Identifier00853541006051
UDI-Public00853541006051
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K181772
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/03/2022
Device Model NumberCTX-A004
Device Catalogue NumberCTX-A004
Device Lot NumberM220034
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer10/19/2022
01/04/2023
10/17/2024
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date10/24/2022
01/05/2023
10/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-