MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problems
Filling Problem (1233); Migration or Expulsion of Device (1395)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/12/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id 8780 lot# serial# (b)(4) product type catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from multiple sources (manufacturer representative, healthcare provider, foreign) regarding a patient who w as receiving an unknown drug of an unknown concentration at an unknown dose rate via an implantable pump for unknown indications for use.It was reported that the patient came for their usual pump refill.The healthcare provider was not able to puncture the refill port.An x-ray was performed which showed that the pump was twisted (flipped).It was noted that the reason was unclear.It was unclear if sutures were used for anchoring.Surgery was scheduled for opening incision and correction of pump positioning.The healthcare provider opened the incision, took out the pump, twisted it to the correct position, and implanted it again.The issue was resolved and there was no further complication.The patient was without injury regarding their status as of (b)(6) 2022.Environmental/external/patient factors that may have led or contributed to the issue was indicated as being not applicable.The pump and catheter remained implanted and in-use.The date of pump implant was unknown.The patient's gender, medical history, age, and weight at the time of the event was unknown or would not be made available.
|
|
Search Alerts/Recalls
|
|
|