Event or Problem Description
|
|
It was reported that on unknown date reporter has some broken frn insertion handles that are needing to be replaced.Intra-operatively sometimes prior to (b)(6) 2022, the threaded portion of two (2) driving caps for insertion handle broke off inside of female end of threaded connection of radiolucent insertion handle for expert nl/175 and radiolucent insertion handle frn.There was a surgical delay of ten (10) minutes.Surgeon had to remove nail, open new set and reassemble insertion handle.The patient and procedure outcome were unknown.This report is for one (1) radiolucent insertion handle for expert nails/175mm.This is report 2 of 4 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part # 03.010.487 lot # 4l07469 manufacturing site: werk hagen dorf release to warehouse date: 17 july 2019 supplier: n/a expiration date: n/a.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|