| Model Number |
DUCT OCCLUDER |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, that a 12-10mm amplatzer duct occluder was successfully implanted in a patient that was previously diagnosed with a patent ductus arteriosus defect and sustained atrial fibrillation.It was noted the following day after implant, that although the patient did not experience any chest symptoms, a "huge negative t-wave" was detected on an electrocardiogram.The patient was experiencing an arrhythmia, but there was no elevation of serum creatine kinase detected.It was noted that the patient's left ventricle apex had become ballooned and there was excessive contraction at the base of the left ventricle.The patient was diagnosed with tako-tsubo-like left ventricular dysfunction.The patient was administered an angiotensin-converting enzyme (ace) inhibitor.A week later, the left ventricle function improved and the negative t-wave detected on the electrocardiogram had normalized.Thereafter, the patient did not experience any symptoms of congestive heart failure for two years.The article concludes that after the closure of the patent ductus arteriosus, the preload of the left ventricle decreases.However in this case study the afterload of the left ventricle increased.Those hemodynamics changes may result in stress to the left ventricle and results in tako-tsubo like ventricular dysfunction.This is especially true in an older patient that has less of an ability to compensate for these hemodynamic changes.This will require more attentive monitoring of geriatric patient's.
|
| |
|
Additional Manufacturer Narrative
|
|
Summarized patient outcomes/complications of amplatzer duct occluder implant were reported in a research article in a subject population with co-morbidity of atrial fibrillation.Some of the complications reported were massive negative t-wave was confirmed, arrhythmia and excessive contraction at the base of the left ventricle; these complications are anticipated for the procedure and subject population.A more comprehensive assessment could not be performed as the event was non-contemporaneously reported through a literature review and no device or individual patient information was received for analysis.
|
| |
|
Search Alerts/Recalls
|