| Model Number |
IPN040971 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that: "the doctor found inflation problem prior to use.".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was then performed and the cuff was inflated to 25mbar using a calibrated endotest meter.No issues were encountered.The sample was then immersed in water for leak testing.No air bubbles were observed coming from the cuff , indicating no leakage.A colored water test was also conducted on the lumen to observe if there is any constriction on side arm fitting that can cause an inflation issue; however, no issues were identified.A device history record review was performed and no relevant findings were identified.The complaint of leakage could not be confirmed.No issues were found with the returned sample.
|
| |
|
Event or Problem Description
|
|
It was reported that: "the doctor found inflation problem prior to use." no patient involvement.
|
| |
|